| Model Number |
SIF-Q180 |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the colonovideoscope to the service center for a separate issue.During device analysis, the service center identified foreign objects in the air/water tube and cylinder.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The final investigation has been completed.The failure identified during the inspection have been confirmed.A definitive root cause could not be identified.However, based on the results of the investigation, the most probable cause of air water tube and cylinder has foreign objects was could not established.Olympus will continue to monitor the field performance of this device and evaluate any additional information received through post market surveillance activities.
|
| |
|