| Model Number |
H749394671500 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Material Separation (1562); Intermittent Loss of Power (4016)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
08/11/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the burr became stuck, detached and was retained in the artery.The 80-90% stenosed target lesion was located in the mildly tortuous and severely calcified ostial circumflex and left main coronary artery.A 1.50mm rotapro and a rotawire were selected for use.During the procedure, it was noted that the burr stalled, became stuck in the lesion and detached.The burr could not be removed from the distal wire.Attempts to retrieve the detached burr via snare were not successful.Th burr was stented in place and the procedure was completed.There were no further patient complications, and the patient was stable after the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Detailed product information was not provided to bsc.Because the product is unknown at this time, we are unable to provide the complete unique identifier (udi) # and other specific product information.If additional details become available, a supplemental report will be submitted.
|
| |
|