| Model Number |
BF-1TQ170 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the videoscope to the service center for evaluation related to a separate issue.During the evaluation, a reportable malfunction was identified: the adhesive on the a-rubber was chipped.As a result, the affected component required replacement.There was no patient involvement associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.Based on the results of the investigation, the probable cause of this complaint is expected or random component failure resulting from separation problem, with no design or manufacturing issue identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.B3: olympus detected the issue during the device servicing; therefore, there is no information regarding the customer reported event date.
|
| |
|