| Model Number |
ASY-13823 |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On (b)(6) 2026 a customer in spain reported to hologic that discrepant results were generated using the aptima combo 2 (ac2) assay (master lot: 930679) with panther instruments sn# (b)(6).On (b)(6) 2026, sample ids (b)(6) were tested in panther worklist id (b)(4) and each generated a chlamydia trachomatis (ct) negative result.The same samples (ids (b)(6)) were retested on another panther instrument on (b)(6) 2026, under worklist id (b)(4), and each generated a chlamydia trachomatis (ct) positive result.No information regarding patient treatment is available at this time.Hologic has not been informed of any adverse patient outcome related to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Hologic reviewed the panther system log files provided by the customer.Review of the logs did not reveal any hardware or instrument issues that contributed to the discrepant result.Following cleaning of the sample/reagent racks and shields with bleach per protocol, no further unexpected results were observed.Aptima lysis tubes processed as a contamination check returned negative results.Hologic has not been informed of any adverse patient outcome related to this event.Hologic further reviewed the impacted master lot 930679 production records which indicated the assay lot was released per qc specifications.
|
| |
|