| Model Number |
G5S-80C |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device prompted a "unit failed" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation.During the investigation it was noted that the reported condition could not be duplicated during depot evaluation; however, review of the device log identified error 500: ecg frontend gain test failure.The main board was replaced to reduce the probability of recurrence and the was recertified for clinical use.Further evaluation of the main board did not determine the root cause.Although the fault could not be reproduced during testing, the log confirms that the critical failure occurred.As a result, the investigation is closed as observed in device data.No trend is associated with reports of this type.
|
| |
|