| Model Number |
H749394671500 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Difficult to Remove (1528); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the burr became stuck on the guidewire.The 90% stenosed target lesion was located in the severely calcified left anterior descending (lad) artery.A 1.50mm rotapro and rotawire drive were selected for use.The rotawire was crossed into the lad and placed across the lesion.When the burr was advanced, it failed to advance on the wire beyond a certain point just proximal to the non-boston scientific guide catheter.When the physician attempted to withdraw the catheter, it would not move.The wire was hydrated, and the rotapro was pulled with force; but could not be withdrawn.Due to the inability to either withdraw or advance the burr, the entire assembly was withdrawn.The procedure was completed using alternate devices.There were no patient complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Good faith effort attempts were made to try and retrieve additional details regarding the reported event, but further information was unable to be obtained.
|
| |
|