| Model Number |
693565 |
| Medical Device Problem Codes |
Fracture (1260); High impedance (1291); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574); Unstable Impedance (4085)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
08/19/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead triggered a lead integrity alert (lia) for high rate non-sustained episodes and sensing integrity counter (sic).The rv lead also exhibited oversensing and varying pacing impedance.The rv lead later triggered a warning for high out of range (oor) pacing impedance.A fracture was confirmed on the rv lead.It was also noted that the patient receive inappropriate high voltage therapy from the rv lead.The rv lead was programmed off and the patient was sent home with a wearable defibrillator vest.The rv lead was later was capped and replaced.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|