The surgical issue questionnaire was reviewed for potential causes of the reported issue.Based on this review, implanting the device after the expiration date, not prepping the surface of the skin with an antiseptic solution, and using incorrect tools have been ruled out.However, a surgical issue was identified as multiple tunneling attempts were made between the two incisions.A curonix representative conducted a review of sterilization and packaging records for the respective product lot; curonix has confirmed that the product was delivered sterile, validated sterilization parameters were used, and sterile barriers were verified to be intact following packaging.The stimulator is used to treat pain.Although the cause of the reported issue is unknown, a surgical issue was identified as multiple tunneling attempts were performed between the two incisions (user error-clinician).Rates reviewed at the most recent complaint trending meeting do not indicate a significant increase in surgical issues.Surgical issue rates remain acceptably low; thus, a capa is not required at this time.Surgical issue rates will continue to be tracked and trended.
|