| Model Number |
TSGC0801 |
| Medical Device Problem Codes |
Leak/Splash (1354); Air/Gas in Device (4062)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/14/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported on 14 august 2026 that the patient presented with mitral regurgitation (tr) for a triclip procedure.During the preparation of triclip steerable guide catheter (tsgc), loss of column was observed at the flush port along with air bubbles.The tsgc was replaced with another device to complete the procedure.There was no clinically significant delay or adverse patient effects reported.
|
| |
|
Additional Manufacturer Narrative
|
|
All available information was investigated and the reported sgc leak/splash (loss of fluid column) and air/gas in device were not confirmed.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history identified no similar complaints from the lot.Based on information provided and the results of the returned device analysis (unable to confirm the reported issues), the cause for the reported leak/splash (loss of fluid column) and air/gas in the device could not be determined.There is no indication of a product issue with respect to manufacture, design or labeling.
|
| |
|