The customer reported that on (b)(6) 2026, the sample port came apart when rn went to flush the catheter; rn disconnected the urine meter tubing from catheter and replaced without issue, and the individual was "fine." medline quality received the reported product in unpackaged, used condition, and an assessment of the returned sample was conducted based on established inspection criteria.Visual assessment revealed that the sample port valve was seated within the housing, and the valve was able to be removed from the housing with little resistance, verifying the reported issue.The reported issue was confirmed with the returned sample and was attributed to an inadequate application of solvent adhesive during manual sample port assembly.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|