• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Allergan (Costa Rica) Style 115 Silicone Gel Filled Breast Implant; prosthesis, breast, noninflatable, internal, silicone gel-filled

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Allergan (Costa Rica) Style 115 Silicone Gel Filled Breast Implant; prosthesis, breast, noninflatable, internal, silicone gel-filled Back to Search Results
Catalog Number 115-290
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/08/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Company representative reported on behalf of patient no complaint against the device.Healthcare professional later reported "intracapsular rupture".Patient later reported through company representative "intracapsular rupture of breast implant", "textured implant included in the recall", "enlarged axillary lymph nodes are likely secondary to systemic process" and "few breast cysts".This record is for the right side.Device was explanted.
 
Additional Manufacturer Narrative
This is a follow-up report to a medwatch submitted under manufacture report number 9617229-2024-11086.A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: observed, broken device assessed as unidentified (tear) opening (shell thickness within specification) and more than three openings were assessed as surgical damage.As per the investigation procedure wear abrasion and creases were completed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Style 115 Silicone Gel Filled Breast Implant
Common Device Name
prosthesis, breast, noninflatable, internal, silicone gel-filled
Manufacturer (Section D)
Allergan (Costa Rica)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
Allergan (Costa Rica)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
+1(847)936-6324
MDR Report Key50675299
Report Number9617229-2026-17117
Device Sequence Number5166595
Product Code FTR
UDI-Device Identifier10888628002371
UDI-Public(01)10888628002371(17)200227
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/27/2020
Device Catalogue Number115-290
Device Lot Number2718973
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer05/27/2024
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/20/2026
Initial Report FDA Received Date09/08/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/27/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age42 YR
Patient SexFemale
Patient Weight123 LB
-
-