| Model Number |
72400024 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
Urinary Incontinence (4572)
|
| Date of Event |
07/01/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient implanted with this artificial urinary sphincter (aus) experienced recurrent incontinence.A surgical procedure was performed, during which fluid loss within the system was identified.The source of the fluid leak was suspected to be the balloon, as this component was found to be devoid of fluid and a hole was observed in the balloon.The aus was completely replaced.No additional patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Model number/catalog number: 72404130.Serial number: n/a.Batch/lot number: 621678016.Model/catalog description: belt cuff fgs 4.0 cm iz.Unique identifier (udi) #: (b)(4).Model number/catalog number: 72404127.Serial number: n/a.Batch/lot number: 644538007.Model/catalog description: control pump fgs iz.Unique identifier (udi) #: (b)(4).
|
| |
|