• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Synthes GmbH 2.7/3.5MM DEPTH GAUGE 0 TO 60MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Synthes GmbH 2.7/3.5MM DEPTH GAUGE 0 TO 60MM Back to Search Results
Catalog Number 03.133.080
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/13/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3:h6 the product was returned to depuy synthes for evaluation, and an investigation was conducted in accordance with depuy synthes procedures.Visual inspection of the returned device found to be broken from the needle.The broken fragment was returned for evaluation.No other anomalies were noted.The reported complaint was confirmed based on the evaluation results.The investigation traced the potential cause to a random component failure that may have been caused by exposure to unintended forces.It has been determined that no corrective or preventive actions are proposed.There is no indication that a design or manufacturing issue caused the reported condition.As part of depuy synthes quality processes, all devices are manufactured, inspected, and released in accordance with approved specifications.Additional monitoring for potential safety signals is conducted through complaint trend analysis as part of post-market safety surveillance activities.Device history lot : part:03.133.080 synthes lot:j013700 supplier lot:j013700 supplier: (b)(6).Release to warehouse date:10 april 2023 expiry date: n/a.No ncrs generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Device history review : part:03.133.080 synthes lot: j013700 supplier lot: j013700 supplier: (b)(6) release to warehouse date:10 april, 2023 expiry date: n/a.No ncrs generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7/3.5MM DEPTH GAUGE 0 TO 60MM
Common Device Name
Depth Gauge
Manufacturer (Section D)
Synthes GmbH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
Monument
1101 synthes avenue
monument CO 80132
Manufacturer Contact
elba bello
1302 wrights lane east
west chester, PA 19380
+1(908)386-3534
MDR Report Key50676565
Report Number8030965-2026-08973
Device Sequence Number17434910
Product Code HTJ
UDI-Device Identifier07612334122032
UDI-Public(01)07612334122032(10)J013700
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
Exempt
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.133.080
Device Lot NumberJ013700
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/14/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/12/2026
Supplement Date Received by Manufacturer09/14/2026
Initial Report FDA Received Date09/08/2026
Supplement Report FDA Received Date09/15/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/12/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-