| Model Number |
TJF-Q180V |
| Medical Device Problem Codes |
Material Discolored (1170); Detachment of Device or Device Component (2907); Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that the duodenovideoscope had a separate issue.During the initial evaluation of the device, discoloration inside the light guide lens, a chip in the adhesive around the light guide lens, and foreign material at the forceps elevator were observed.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event is unknown.Additional information will be provided if available.Based on the final investigation of the device the initial evaluation findings were confirmed.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: the most probable cause of the discoloration inside the light guide lens and the chip in the adhesive around the light guide lens was traced to a component failure.However, the most probable cause of the foreign material at the forceps elevator could not be established.The legal manufacture reviewed the customer provided the cleaning disinfection and sterilization (cds) processes where the following deviation from instructions for use (ifu) was confirmed: the customer did not clean, disinfect, and sterilize the device before sending it to olympus.Olympus will continue to monitor the field performance of this device.Should any additional relevant information become available, a supplemental report will be submitted.
|
| |
|