| Model Number |
AIA-360 IMMUNOASSAY ANALYZER |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/14/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported failure mode ¿2012 air detected (diluent)¿ on the aia-360 analyzer.The customer primed the analyzer and no air was noted in the lines.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event which resulted in a delayed reporting of patient samples for beta human chorionic gonadotropin (bhcg).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|
| |
|
Additional Manufacturer Narrative
|
|
A field service engineer (fse) was dispatched to address the reported event.The fse confirmed the error during troubleshooting and inspected the analyzer.Upon opening the analyzer cover, the fse identified a bent sample nozzle.The fse replaced the sample nozzle and determined that some sample tube holders were deformed, which had caused the sample needle to bend.The fse validated the analyzer by successfully performing daily check and quality control without error, and within acceptable range.No further action required by field service.The aia-360 analyzer is functioning as expected.The aia-360 operator's manual under section7-1: error messages states the following:
(2012) air detected "[diluent]".Description: there is no contact with the liquid after diluent suction.Troubleshooting: contact the service department.The most probable cause of the reported event was due to deformation of the sample tube holders.
|
| |
|
Search Alerts/Recalls
|