| Model Number |
IO-401 |
| Medical Device Problem Code |
Insufficient Flow or Under Infusion (2182)
|
| Health Effect - Clinical Codes |
Dyspnea (1816); Low Oxygen Saturation (2477)
|
| Date of Event |
08/17/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
B3 date of event is estimated.The unit was returned for evaluation on (b)(6) 2026.The unit received a reported oxygen error, which was confirmed through the data log that shows sieve warning.Internal low (80.7%), psa (9) and external 02 reading servomex (79.7%).The issue was caused by contaminated zeolite, indicating that the zeolite absorbed excessive moisture.Filter frames are dirty due to dust.Housing was scratched probably due to rough cleaning.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's unit had a system error message.As a result, the unit was not producing enough oxygen resulting in shortness of breath.A replacement unit was sent to the patient.The inogen team attempted to contact the patient for further information three times with no response.
|
| |
|