BOSTON SCIENTIFIC CORPORATION Polaris? Ultra; stent, ureteral Back to Search Results
Model Number M0061921230
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/12/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a polaris ultra ureteral stent was to be used in a lithotripsy procedure.During unpacking, it was found that the stent was fractured.The procedure was completed with another of the same device and there were no patient complications reported.
 
Event or Problem Description
It was reported that a polaris ultra ureteral stent was to be used in a lithotripsy procedure.During unpacking, it was found that the stent was fractured.The procedure was completed with another of the same device and there were no patient complications reported.
 
Additional Manufacturer Narrative
Block h6: device code a0401 captures the reportable event of stent shaft break.
 
Additional Manufacturer Narrative
Block h6: device code a0401 captures the reportable event of stent shaft break.Correction to block h6 device code.Block h11: investigation analysis the device was not returned for analysis; therefore, a technical analysis could not be performed.A review of the device history record (dhr) was performed.It was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.A risk review was completed and confirmed that the event of stent shaft break and coil detachment of device or device component was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.A similar complaint review was completed.There were no emerging trends identified that warrant further action.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Polaris? Ultra
Common Device Name
stent, ureteral
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
2546 calle primera
alajuela
CS  
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50684087
Report Number2124215-2026-48076
Device Sequence Number5155688
Product Code FAD
UDI-Device Identifier08714729124320
UDI-Public(01)08714729124320(17)271002(10)34979139
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K010002
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM0061921230
Device Catalogue NumberM0061921230
Device Lot Number0034979139
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/14/2026
Supplement Date Received by Manufacturer09/07/2026
Initial Report FDA Received Date09/09/2026
Supplement Report FDA Received Date09/30/2026
Date Device Manufactured10/02/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age50 YR
Patient SexMale
Patient Weight154 LB
-
-