• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medtronic Xomed Inc. XPS® HANDPIECE - SKEETER® OTOLOGIC DRILL; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) I

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medtronic Xomed Inc. XPS® HANDPIECE - SKEETER® OTOLOGIC DRILL; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) I Back to Search Results
Model Number 3055601
Medical Device Problem Code Vibration (1674)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/14/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during use the blade wobbles, have an oscillation during rotation. the user informs that several blades of various batches have been tried and the problem remains.There was no patient impact.
 
Additional Manufacturer Narrative
H3: analysis of the sample is not completed as of the date of this report, a follow-up report will be submitted once analysis gets c ompleted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS® HANDPIECE - SKEETER® OTOLOGIC DRILL
Common Device Name
DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) I
Manufacturer (Section D)
Medtronic Xomed Inc.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
Medtronic Xomed Inc.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
justin ellis
6743 southpoint drive north
jacksonville, FL 32216
+1(763)526-5677
MDR Report Key50684091
Report Number1045254-2026-01243
Device Sequence Number5315750
Product Code ERL
UDI-Device Identifier00763000378004
UDI-Public(01)00763000378004(11)250707
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K833874
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3055601
Device Catalogue Number3055601
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/20/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2026
Supplement Date Received by Manufacturer09/10/2026
Initial Report FDA Received Date09/09/2026
Supplement Report FDA Received Date09/16/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/07/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-