| Model Number |
Cardiohelp-i |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that emergency drive was broken during use however customer does not want to provide demographics.Later it was found that the latch of the emergency drive was sticky and therefore was not moving.No harm to any person has been reported.As there is a potential risk, that in emergency situations the patient cannot be supported via the hls set and emergency drive till a new cardiohelp is connected, a report is required.Complaint # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
A getinge technician was sent for investigation on 2026-08-28.The technician confirmed that no damage was found on the emergency drive however latch not retracting, it was sticky and not moving freely.Cleaning solved the problem.No parts were replaced.The technician performed safety, calibration, and functionality checks to factory specifications.All function tests are passed.A follow up medwatch will be submitted when further information becomes available.
|
| |
|