| Model Number |
0998-00-0800-31 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the customer that during standby, while charging, the cardiosave intra-aortic balloon pump (iabp) was not turning on.There was no patient involvement.No harm was reported.An issue was identified with the 9v battery - it has corroded onto the alarm daughter board.Alarm daughter board and alarm board to exec processor cable have been ordered for replacement.
|
| |
|
Additional Manufacturer Narrative
|
|
E1 - event site name: (b)(6) hospital.Upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|