The customer reported that when powering on, turns off by itself, and only powers on with the power cord.Per customer, "no error messages are displayed." there was no patient impact or consequence reported.
|
|
Other, other text: the returned device was evaluated.During the evaluation, the device powered on using both battery and ac power.However, after displaying the splash screen, the display went blank, followed by an alarm state consisting of seven beeps.During evaluation it was determined a defective speaker prevented the device from completing the power-on sequence and generating the alarm.Additionally, an ¿audio speaker fault" error message was identified in the device log files.The reported issue was confirmed during testing.Health effect impact code: no adverse event; no patient impact.
|