| Catalog Number |
CF7-7-60 |
| Medical Device Problem Codes |
Overheating of Device (1437); Deformation Due to Compressive Stress (2889)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/02/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a closurefast catheter was being used to treat the great saphenous vein (gsv).Tumescent infiltration, local anesthesia and hand compression were utilized. a guidewire was used for the insertion of the catheter.The proper temperature was reached.During the procedure the catheter became kinked on the coil.The catheter was replaced with a new one, and the procedure completed, two segments were treated and the vein closed. the lumen was flushed prior to use and the ifu (instruction for use) was followed during preparation, procedure, post-procedure. no patient complications have been reported as a result of this event.When the device was returned for evaluation, it was noted that the catheter was damaged.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: damage was noted to the catheter heating element/coil.Examination revealed bubbling/bending/kinking of the heating element/coil.The heating element/coil was not exposed; white residue was observed beneath the fep layer and on the element coils.The bubbling/burning/bending/kinking was observed at approx.1.0 cm, 7.0 cm from the distal tip of the device.The catheter cable connector was examined: the catheter reference key shows evidence of wear, and the red ink was visible in some areas.All pins were visually inspected to be straight.Visual inspection of the returned catheter revealed three kinks along the shaft; the kinks were observed at approx.14.2 cm, 42.9 cm, 49.1 cm from the distal tip of the device.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|