| Model Number |
N/A |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Thrombosis/Thrombus (4440)
|
| Date of Event |
08/12/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Event reported as thrombus
distal thrombus inside the stent at the junction of the graft.One day following the initial procedure, the patient was placed on aspirin and had a ct(computed tomography) scan with the vascular surgeons as part of a follow-up for tevar(thoracic endovascular aortic repair).Thombus identified on transoesophageal echocardiography (toe).The treating vascular surgeon reported that the thoracic stent graft was appropriately deployed and fully expanded.6hr bypass time, blood transfusion given but no excessive products given.No death/serious injury.No surgical intervention.Sales associate has requested patient condition follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Clinical code
4440 - thrombosis/thrombus: event reported as thrombosis.Impact code
4644 - medication required - the site reported that the patient was placed on aspirin.The stent graft was reported by the treating surgeon to be appropriately deployed and fully expanded.Medical device problem
3190 - insufficient device problem information: awaiting further information.Component code
4755 - part/component/sub-assembly term not applicable.Type of investigation
4109 - historical data analysis: four-year review was performed for thoraflex hybrid > occlusion/thrombosis > body/unknown/blank jan22 - aug26 which gave an occurrence score of 0.116%.Investigation finding
3233 - results pending completion of investigation.Investigation conclusion
11 - conclusion not yet available.
|
| |
|
Search Alerts/Recalls
|