| Model Number |
8637-40 |
| Medical Device Problem Code |
Electromagnetic Interference (1194)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received from multiple sources (healthcare provider, clinical study) regarding a patient who was receiving unknown drug via an implantable pump for unknown indications for use.It was reported that the patient had a pump motor stall today.They are thinking it might be from her iphone and due to a magnetic issue.Additional information received from the healthcare provider via a clinical study indicated that the motor stall resolved spontaneously.They believed that possibly the apple iphone cover had a magnetic reaction.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|