The customer reported concern about the pregnancy test because ¿she has got three faintly positive,¿ and additional information stated the reason for the test was ¿prior to iud insertion¿; no adverse patient, fetus, or user event, no delay in a medical procedure or treatment, no treatment/procedure provided or withheld based on the result, ¿none¿ for impact, and no adverse patient outcomes were reported.Complaint could not be confirmed as there were no samples or pictures received; retained devices from the reported lot number were tested with hcg-negative clinical urine samples, the results were read at 3 and 4 minutes, and all devices yielded the expected negative results with no false positive results observed during in-house testing.The reported issue was not replicated during testing of retention product, and a root cause could not be determined from the available information.The case details were reviewed along with the complaint history for the reported issue and no indications of a systemic issue were identified; manufacturing batch record review did not uncover any relevant non-conformances and found that the lot met quality control specifications.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|