| Model Number |
D143 |
| Medical Device Problem Codes |
Over-Sensing (1438); Failure to Read Input Signal (1581)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited far-field p-wave oversensing on the right ventricular (rv) channel.It was also noted there was an rv amplitude that was out of range.Technical services (ts) reviewed the data and noted the r-wave was marked as ventricular fibrillation (vf) and there was no blanking for the atrial sense.Ts recommended performing a defibrillation threshold (dft) test and suggested reprogramming options.No adverse patient effects were reported.This icd remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|