| Model Number |
D142 |
| Medical Device Problem Codes |
Over-Sensing (1438); Pacing Problem (1439)
|
| Health Effect - Clinical Code |
Syncope/Fainting (4411)
|
| Date of Event |
08/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient with this implantable cardioverter defibrillator (icd) presented to the emergency room (er) due to syncope.The health care professional (hcp) noticed pauses on the electrograms (egms) and requested a review of reports from technical services (ts).Ts reviewed the data and noticed there were some pacemaker mediated tachycardia (pmt) episodes which showed dropped beats along with intermittent t-wave oversensing (twos).It was noted the twos was also causing dropped beats.Ts discussed the patient might be sensitive to the dropped beats and recommended reprogramming options.The device was reprogrammed and no additional adverse patient effects were reported.This icd remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|