| Model Number |
D120 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited high out of range pace impedance measurements which subsequently dropped.Technical services (ts) was consulted and found a non-boston scientific right ventricular lead was implanted which had an old header.Ts noted the caller denied non-physiological noise and recommended troubleshooting with isometrics and pocket manipulation.No adverse patient effects were reported.This icd remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|