Model Number GS-100
Medical Device Problem Codes Degraded (1153); Particulates (1451); Improper Flow or Infusion (2954); Noise, Audible (3273)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/22/2026
Type of Reportable Event Malfunction
Event or Problem Description
Device: inogen at home oxygen concentrator prescribed continuous oxygen at flow setting 2 around the clock and setting 4 overnight with bipap (resmed aircurve 10 st-a).This serial was shipped (b)(6) 2025 as a new warranty advance replacement after refurbished sn (b)(6) failed out of the box (delivered (b)(6) 2025, returned (b)(6) 2025).Prior original (purchased (b)(6) 2022) had also failed with a level-4 rattle and was replaced.On the night of (b)(6) 2026 the unit developed a sudden abnormal mechanical noise.The noise is audible at setting 2 (continuous therapy) and louder at setting 4 (overnight bipap).It increases with flow.Particle filter was clean and intake unobstructed (video).Videos of level 2, level 4, and filter were sent to inogen.This unit exhibited a sudden, flow-dependent mechanical noise at 698 hours -- about 3% of published compressor life -- with a clean filter.That pattern is consistent with premature compressor/mechanical failure, not user maintenance error.Around (b)(6) 2026, while operating at setting 2, the front panel showed solid yellow warning light, flow indicator #4, and solid blue light, with no beep noticed.Lights were gone the next morning after restart; -28- remains the stored event.Low oxygen is low-priority and does not itself prove shutdown, but it does prove a recorded purity event.On (b)(6) 2026, tilting the running unit forward produced white "smoke"/mist from the rear bottom vent during part of the cycle.Not visible (or too faint) when upright.Dark-cloth residue check at the vent: dry white powder, no moisture.That is consistent with zeolite (sieve) dust in the exhaust path, not condensate and not a fire-hazard alarm (-9- was not shown).Product is still in the home.Patient is using a medicare-provided concentrator as backup.A replacement oxyhome 5l was purchased separately.Inogen dispute abandoned; this report is for fda device-safety surveillance, not a warranty claim.Columns flagged o2 service soon and low oxygen far short of 18¿24 months; sudden flow-dependent mechanical noise after nine quiet months; clean particle filter on video; dry white powder at rear exhaust when tilted (consistent with sieve-media dust).Combined, this is premature mechanical/sieve failure of a 2025-shipped unit, not end-of-life wear.Fda is asked to record the malfunction and the postmarket service refusal on a life-supporting home oxygen device.Reference reports mw6002022, mw6002023.Annex a: 3273, 1451, 1153, 2954.Annex e: 4582.Annex f: 2199.Annex g: 4755.This report captures inogen at home oxygen concentrator # 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Inogen At Home Oxygen Concentrator
Common Device Name
Generator, Oxygen, Portable
Manufacturer (Section D)
Inogen, Inc.
MDR Report Key50690648
Report NumberMW6002021
Device Sequence Number5915565
Product Code CAW
UDI-Device Identifier00817131020148
UDI-Public(01)00817131020148
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Voluntary
Initial Reporter Occupation Patient
Type of Report Initial
Report Date (Section B) 08/31/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberGS-100
Device Catalogue Number10-06316-00-01
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date09/09/2026
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
Patient Weight88 LB
Patient RaceWhite
-
-