|
It was reported that the patient presented with grade ii mobility of an is bridge.Radiographic examination revealed significant bone loss.Clinical examination showed exposure of the implant threads.The patient was diagnosed with peri-implantitis.It was decided to remove the implants in order to perform guided bone regeneration (gbr) and subsequently place new implants.Tooth sites 11, 13, 22 and 23.
|
|
Zimvie complaint number (b)(4).A2: age at time of the event: unknown / not provided.A4: patient weight: unknown / not provided.D10: concomitant medical products:.Tsvb13, impl tapered scr-v sbm 3.7mm 3.5mm 13mm, lot: 1266763.Tsx37b11, tsx implant, 3.7mmd, 11.5mml, lot: 1271303.Tsx37b10, tsx implant, 3.7mmd, 10mml, lott: 1272875.E1: address-line 2: calle 62a.E1: phone #: unknown / not provided.G4: premarket identification pma/510(k) #: k011028|k013227.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|