BOSTON SCIENTIFIC CORPORATION RESONATE? EL ICD DR; Implantable Cardioverter Defibrillator (ICD) Back to Search Results
Model Number D433
Medical Device Problem Code Over-Sensing (1438)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/08/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a hospital visit, the health care professional (hcp) called technical services (ts) for a consult review of this implantable cardioverter defibrillator (icd) presenting electrogram (egm).Upon review, right atrial (ra) far field oversensing which led to inappropriate atrial tachy response (atr) was observed.Ts provided the hcp with programming recommendations.At this time, the icd remains in service.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONATE? EL ICD DR
Common Device Name
Implantable Cardioverter Defibrillator (ICD)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50691012
Report Number2124215-2026-46733
Device Sequence Number14934329
Product Code LWS
UDI-Device Identifier00802526588075
UDI-Public(01)00802526588075(17)270620
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P960040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberD433
Device Catalogue NumberD433
Device Lot Number707905
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2026
Initial Report FDA Received Date09/09/2026
Date Device Manufactured06/29/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age46 YR
Patient SexFemale
-
-