|
The customer reported that the coagulation part of the pen was fine, but the cutting part would not work at all.Medline quality received 1 ea for evaluation, and a functional inspection was conducted with the provided sample; the sample was plugged into a bovie generator and functioned properly when using cut mode.The reported concern could not be confirmed, as the sample functioned normally on cutting mode, and the cause was documented as not confirmed with received sample.A retrospective review of the past year indicated that there have been no prior complaints in regard to this issue, and the division will continue to monitor this issue for trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|