ZimVie US Corp LLC IMP,TSV,4.1MM,SBM,13; Dental Implant Back to Search Results
Catalog Number TSV4B13
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Osteolysis (2377)
Date of Event 07/18/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported significant bone loss due to peri implantitis, tooth sites 13 and 14.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4) a2: age at time of the event: unknown / not provided.A4: patient weight: unknown / not provided.D10: concomitant medical products:.Tsvb13, impl tapered scr-v sbm 3.7mm 3.5mm 13mm, lot: 1258962.E1: address-line 2: calle 62a.E1: phone #: unknown / not provided.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMP,TSV,4.1MM,SBM,13
Common Device Name
Dental Implant
Manufacturer (Section D)
ZimVie US Corp LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZimVie US Corp LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
mark mashburn
4555 riverside drive
palm beach gardens, FL 33410
+1(561)971-3230
MDR Report Key50691736
Report Number0001038806-2026-04731
Device Sequence Number5891753
Product Code DZE
UDI-Device Identifier00889024019249
UDI-Public(01)00889024019249(17)280109(10)1261281
Combination Product (Y/N)N
Initial Reporter CountryCS
PMA/510(K) Number
K072589
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberTSV4B13
Device Lot Number1261281
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 08/28/2026
Initial Report FDA Received Date09/09/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/10/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Dental Implant See H11 Narrative
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient SexFemale
Patient Weight198 LB
-
-