The reported issue was acsd luer lock access device not working properly; they aren¿t pulling blood like they should, and using a syringe to aspirate the needle failed as well.Medline received 1cs, acsd luer lock access device in new condition, randomly selected 20ea sample for testing, and performed functional testing utilizing a venipucture blood board to simulate blood sample collection.Medline was able to draw the "blood" with no issues, saw no indications of leakage coming from the needle rubber cover or the rubber cover not coming back to the initial position exposing the needle, and the complaint was not confirmed.The cause was not confirmed with received sample under unconfirmed defect; the samples worked as intended, and medline will continue to monitor this issue for trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|