The customer reported receiving a medline safety needle insulin 0.5ml that didn¿t successfully activate safety mechanism and reported that a nurse got stuck by the needle because the safety part didn't stay closed on the needle right.The sample(s) were marked received by medline on (b)(6) 2026, and storage, mailing room, and various labs were periodically checked for this sample.After 5 months the sample still could not be located, the complaint was marked not confirmed, and medline was not able to investigate it.Sample was required to determine cause; no trends were identified for the specified material number and its corresponding defect type, and there has been no change to the reportability decision associated with this complaint.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|