| Catalog Number |
EVO-FRM-01 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Bruise/Contusion (1754); Obstruction/Occlusion (2422); Pallor (2468)
|
| Date of Event |
08/21/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
An hcp reported injecting 0.2 cc of evolysse form into a patient's pyriform aperture using a 25-gauge cannula.When the tissue blanched, the hcp immediately withdrew the cannula, injected two vials of hylenex, applied a warm compress, and massaged the area.The patient subsequently developed bruising and mottling.An md confirmed a vascular occlusion in the left nasolabial fold, which resolved after ultrasound-guided hylenex treatment.
|
| |
|
Event or Problem Description
|
|
An hcp reported injecting (b)(6)of evolysse form into a patient's pyriform aperture using a 25-gauge cannula.When the tissue blanched, the hcp immediately withdrew the cannula, injected two vials of hylenex, applied a warm compress, and massaged the area.The patient subsequently developed bruising and mottling.An md confirmed a vascular occlusion in the left nasolabial fold, which resolved after ultrasound-guided hylenex treatment.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was injected into the patient and is not available for return.Further information regarding event, product, or patient details has been requested.No additional information is available at this time.Based on the reported information, this event is consistent with a vascular occlusion which is a known serious adverse event documented in the labeling.This can occur due to unintended injection into a blood vessel or injection of a large amount of filler near a blood vessel causing compression of the blood vessel.Both occlusion and compression may result in reduced or cessation of blood supply to tissues.Intervention is required to prevent serious and permanent complications.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was injected into the patient and is not available for return.Further information regarding event, product, or patient details has been requested.No additional information is available at this time.Based on the reported information, this event is consistent with a vascular occlusion which is a known serious adverse event documented in the labeling.This can occur due to unintended injection into a blood vessel or injection of a large amount of filler near a blood vessel causing compression of the blood vessel.Both occlusion and compression may result in reduced or cessation of blood supply to tissues.Intervention is required to prevent serious and permanent complications.Follow-up 1: 17-sep-2026
the batch release form was reviewed, and the device met all specifications prior to release.D4 additional device information: lot number, udi, and expiration date were added
(b)(4).
|
| |
|
Search Alerts/Recalls
|