| Model Number |
03.58.263 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Bone Fracture(s) (1870)
|
| Date of Event |
08/18/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Spinal fixation was performed at l4-l5.The l5 pedicle fractured after the rod was inserted.The trajectory was changed from cbt to ps, and the same screw was reinserted.As the trajectory was altered, the required rod length changed, and a new rod (50 mm) was used.No critical delay.The quality of the patient`s bone was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Batch review performed on 20 aug 2026 lot 2228100: (b)(4) items manufactured and released on 09-dec-2022.Expiration date: 2027-nov-21.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Root cause: although no specific root cause can be confirmed, the issue experienced is likely related to the unique patient conditions and surgical application.There is no indication that any potential issue with the device may have caused or contributed to the event, and the device history record review does not indicate any potential manufacturing related issue.
|
| |
|