Model Number 01.12.143
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Medical device site infection (4845)
Date of Event 08/25/2026
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2022, the patient had primary surgery.On (b)(6) 2026, the patient came in due to infection and the pathogen is unknown.The surgeon performed a washout, revised the 28mm biolox head s to a same size head and revised the d 28/dmi hc liner to a same size liner.The surgery was completed successfully.Presently, on (b)(4) 2026, the patient came in due to signs of infection and the pathogen is unknown.The surgeon performed a washout, i&d, and revised all implants to antibiotic spacers.The surgery was completed successfully.
 
Additional Manufacturer Narrative
Batch review performed on 27 aug 2026 stem: quadra-p 01.12.143 quadra-p lat.Size 3 (k181254) lot.2409877: (b)(4) items manufactured and released on 03/jul/2024.Expiration date: 20/jun/2029.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Mpact dm 01.26.2858mhc double mobility hc liner d 28/dmi (k092265) lot.2526503: (b)(4) items manufactured and released on 16/dec/2025.Expiration date: 24/nov/2030.No anomalies found related to the problem.To date, (b)(4) of the same lot have been sold with no similar reported event during the period of review.Ball heads: mectacer 01.29.201 mectacer head biolox delta dia.28 12/14-s (k112115) lot.2604841: (b)(4) items manufactured and released on 18/mar/2026.Expiration date: 03/mar/2031.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Cup: mpact 01.32.158mb mpact dm acetabular shell d 58 (k143453) lot.2427262: (b)(4) items manufactured and released on 05/may/2025.Expiration date: 15/apr/2030.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Root cause: infection is a known possible complication of any surgery.Although no root cause can be established, there is no indication that any issue with the device may have caused or contributed to the event, and the document review does not indicate any potential manufacturing related cause.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Quadra P femoral stems
Common Device Name
Quadra-P Lat. size3
Manufacturer (Section D)
Medacta International SA
strada regina
castel san pietro 6874
SZ  6874
Manufacturer Contact
marco giannessi
strada regina
castel san pietro 6874
SZ   6874
MDR Report Key50692295
Report Number3005180920-2026-00944
Device Sequence Number9685016
Product Code LZO
UDI-Device Identifier07630040720557
UDI-Public07630040720557
Combination Product (Y/N)N
PMA/510(K) Number
K181254
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number01.12.143
Device Lot Number2409877
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2026
Initial Report FDA Received Date09/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/03/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexMale
-
-