| Model Number |
01.29.209 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Joint Dislocation (2374); Implant Pain (4561)
|
| Date of Event |
08/23/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
At about 3 days after the primary, the patient came in presenting pain due to dislocation of the head from the liner and the cause is unknown.The surgeon revised theflat pe hc linerd 36/eto a hooded pe hc liner d 36/e and themectacer head biolox delta d 36 12/14-mto a biolox head l.The surgery was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Batch review performed on 26 august 2026 ball heads: mectacer 01.29.209 mectacer head biolox delta dia.36 12/14-m (k112115) lot 2614373: (b)(4) items manufactured and released on 10-jun-2026.Expiration date: 21-may-2031.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Liner: mpact 01.32.3644hct flat pe hc liner d 36/e (k103721) lot 2602343: (b)(4) items manufactured and released on 06-may-2026.Expiration date: 19-apr-2031.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Root cause: implant dislocation is possible literature described adverse event after primary joint arthroplasties and causes are often unknown.There is no indication that any potential issue with the device may have caused or contributed to the event and the investigation does not indicate any potential manufacturing related issue or systemic deficiency.
|
| |
|