Model Number 04.01.0170
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 08/25/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery due to a dislocation of the liner and glenosphere after about 2 years and 3 months from primary.The surgeon revised the +0mm/0° metaphysis to a same size metaphysis, revised the d 39/+3mm reverse liner to a d 42/+0 constrained liner, and revised the d 39x24.5 glenosphere to a 42xd 24.5 lat.Glenosphere.
 
Additional Manufacturer Narrative
Batch review performed on 09 september 2026 reverse shoulder system 04.01.0170 glenosphere - d 39x24.5 (k170452) lot.2339652: (b)(4) items manufactured and released on 23 january 2024.Expiration date: 07 january 2029.No anomalies found related to the problem.To date, all items of the same lot have been sold, with no similar reported events during the period of review.Reverse shoulder system 04.01.0123 humeral reverse hc liner d 39/+3mm (k170452) lot.2315954: (b)(4) items manufactured and released on 24 august 2023.Expiration date: 02 august 2028.No anomalies found related to the problem.To date, all items of the same lot have been sold, with no similar reported events during the period of review.Root cause: dislocation is possible literature described adverse event after primary joint arthroplasties and causes are often unknown.There is no indication that any potential issue with the device may have caused or contributed to the event and the investigation does not indicate any potential manufacturing related issue or systemic deficiency.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Shoulder System
Common Device Name
Glenosphere - Ø39x24.5
Manufacturer (Section D)
Medacta International SA
strada regina
castel san pietro 6874
SZ  6874
Manufacturer Contact
marco giannessi
strada regina
castel san pietro 6874
SZ   6874
MDR Report Key50692305
Report Number3005180920-2026-00926
Device Sequence Number18540548
Product Code PHX
UDI-Device Identifier07630040706575
UDI-Public07630040706575
Combination Product (Y/N)N
PMA/510(K) Number
K170452
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number04.01.0170
Device Lot Number2339652
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2026
Initial Report FDA Received Date09/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/23/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexMale
-
-