| Model Number |
01.32.152MB |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Joint Dislocation (2374); Implant Pain (4561)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
At about 2 months from the primary, the patient came in presenting pain due to a dislocation of the liner from the cup and the cause is unknown.The surgeon revised the d 28/dmf liner to a same size liner and revised the 28mm biolox delta head s to a cocr ball head 12/14 d 28 mm size xl.The surgery was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Batch review performed on (b)(6) 2026 cup: mpact 01.32.152mb mpact dm acetabular shell d 52 (k143453) lot 2532982: (b)(4) items manufactured and released on 16-mar-2026.Expiration date: 02-mar-2031.No anomalies found related to the problem.To date, 71 items of the same lot have been sold with no similar reported event during the period of review.Liner: mpact dm 01.26.2852mhc double mobility hc liner d 28/dmf (k092265) lot 2530203: (b)(4) items manufactured and released on 29-jan-2026.Expiration date: 23-dec-2030.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Root cause: implant dislocation is possible literature described adverse event after primary joint arthroplasties and causes are often unknown.There is no indication that any potential issue with the device may have caused or contributed to the event and the investigation does not indicate any potential manufacturing related issue or systemic deficiency.
|
| |
|