| Model Number |
01.29.204 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Joint Dislocation (2374); Implant Pain (4561)
|
| Date of Event |
08/28/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The patient had pain due to joint luxation; the head came off the liner, and the cause is unknown.There was no indication of trauma or fall.At about 1 month and a half post-primary.The surgeon revised the two-hole acetabular shell to an mpact dm acetabular shell d 50, revised the 32mm biolox head s to a 28mm biolox head l, and revised the flat pe hc liner d 32/d to a double mobility hc liner d 28/dme.The surgery was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Batch reviews performed on 31 august 2026: ball heads: mectacer 01.29.204 mectacer head biolox delta dia.32 12/14-s (k112115) lot 2529444: (b)(4) items manufactured and released on 28-nov-2025.Expiration date: 16-nov-2030.No anomalies were found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Liner: mpact 01.32.3241hct flat pe hc liner d 32/d (k103721) lot 2530220: (b)(4) items manufactured and released on 17-dec-2025.Expiration date: 26-nov-2030.No anomalies were found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Root cause: based on the information available, no definitive root cause can be established.No trauma or fall was reported, and the batch review did not identify any potentially related manufacturing issue.There is no indication that a potential device-related issue may have caused or contributed to the event.
|
| |
|
Search Alerts/Recalls
|