| Model Number |
01.18.133 |
| Medical Device Problem Code |
Osseointegration Problem (3003)
|
| Health Effect - Clinical Codes |
Inadequate Osseointegration (2646); Aseptic loosening (4883)
|
| Date of Event |
08/19/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Revision surgery due to stem loosening after about 13 years and 4 months from primary.The surgeon revised the medacta stem and head to a competitor stem and head, and revised the head biolox delta dia.28 12/14-l to a double mobility hc liner d 28/dmg.
|
| |
|
Additional Manufacturer Narrative
|
|
Batch review performed on 04 september 2026 lot.125014: 30 items manufactured and released on 13 february 2013.Expiration date: 31 december 2017.No anomalies found related to the problem.To date, all items of the same lot have been sold, with no similar reported events during the period of review.Root cause:aseptic loosening is a known possible complication of any surgery.Although no root cause can be established, there is no indication that any issue with the device may have caused or contributed to the event, and the document review does not indicate any potential manufacturing related cause.
|
| |
|