| Model Number |
04.01.0119 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
08/21/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
At about 1 month from the primary, the patient came in presenting pain due to dislocation of the liner from the glenosphere while making bed.The surgeon revised the d 36x24.5 glenosphere to a d 39x24.5 glenosphere, the metaphysis to a same size metaphysis and the d 36/+0mm liner to a d 39/+5mm liner.The surgery was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Batch review performed on 24 august 2026 reverse shoulder system 04.01.0119 humeral reverse hc liner d 36/+0mm (k170452) lot 2524252: 90 items manufactured and released on 05-nov-2025.Expiration date: 26-oct-2030.No anomalies found related to the problem.To date, (b)(4)items of the same lot have been sold with no similar reported event during the period of review.Reverse shoulder system 04.01.0169 glenosphere - d 36x24.5 (k170452) lot 2525649: (b)(4)items manufactured and released on 07-jan-2026.Expiration date: 14-dec-2030.No anomalies found related to the problem.To date, all items of the same lot have been sold with no similar reported event during the period of review.Root cause:implant dislocation is possible literature described adverse event after primary joint arthroplasties and causes are often unknown.There is no indication that any potential issue with the device may have caused or contributed to the event and the investigation does not indicate any potential manufacturing related issue or systemic deficiency.
|
| |
|