| Model Number |
01.29.230H |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Medical device site infection (4845)
|
| Date of Event |
08/26/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The patient had primary right hip surgery on (b)(6)2026.On (b)(6)2026, the patient came in due to signs of infection and the pathogen is unknown.The surgeon performed a washout and revised the 28mm biolox delta head m to a 28mm biolox option head with option sleeve m, and revised the double mobility hc liner d 28/dmg to a same size liner.The surgery was completed successfully.Presently, the patient came in due to continued signs of infection and the pathogen is unknown.The surgeon revised all components (stem, head, liner, cup) to antibiotic spacers until the infection clears.The surgery was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Batch review performed on 31 august 2026 ball heads: mectacer 01.29.230h mectacer head biolox option 12/14 d 28 (k131518) lot 2606604: (b)(4)items manufactured and released on 07-apr-2026.Expiration date: 16-mar-2031.No anomalies found related to the problem.To date, (b)(4)items of the same lot have been sold with no similar reported event during the period of review.Cup: mpact 01.32.154mb mpact dm acetabular shell d 54 (k143453) lot 2518723: 60 items manufactured and released on 11-feb-2026.Expiration date: 02-feb-2031.No anomalies found related to the problem.To date, (b)(4)items of the same lot have been sold with no similar reported event during the period of review.Liner: mpact dm 01.26.2854mhc double mobility hc liner d 28/dmg (k092265) lot 2601012: (b)(4)items manufactured and released on 2-mar-2026.Expiration date: 11-feb-2031.No anomalies found related to the problem.To date, 64 items of the same lot have been sold with no similar reported event during the period of review.Stem: amistem-p collared 01.18.444 amistem-p collared lat.Size 4 (k173794) lot 2300481: (b)(4) items manufactured and released on 2-may-2023.Expiration date: 5-apr-2028.No anomalies found related to the problem.To date, (b)(4)items of the same lot have been sold with no similar reported event during the period of review.Root cause:infection is a known possible complication of any surgery.Although no root cause can be established, there is no indication that any issue with the device may have caused or contributed to the event, and the document review does not indicate any potential manufacturing related cause.
|
| |
|
Search Alerts/Recalls
|