Orthofix was made aware on 12 aug 2026 that a patient underwent spinal surgery consisting of the shoreline acs system.The index surgery occurred on (b)(6) 2026.It was reported that an 85-2107 7mm, locking cover fractured into the interbody intra-operatively and the threaded portion was left in the interbody.The fracture occurred when the surgeon tried to advance the 85-2107 7mm, locking cover using the locking handle.According to the reporter, the 85-2107 7mm, locking cover did not advance and was not seated properly.Since the interbody was already fixated at c6-7 with the screws, the surgeon elected to leave the construct in place and not remove the fractured threaded portion of the locking cover.No patient injuries or additional medical intervention were reported.Per the reporter, the patient is recovering well.
|
|
The affected device was not returned for evaluation.The lot number was not provided; therefore, a device history record review could not be performed.According to the reporter, the 85-2107 7mm, locking cover was not available for evaluation because it was discarded.A photo was provided which appears to confirm the reported failure mode that the threaded portion of the locking cover fractured.Due to the absence of the returned device the specific root cause cannot be conclusively determined.No patient injury, revision surgery, or additional medical intervention has been reported.Review of labeling:
possible adverse events:
like other spinal system implants, the following adverse events are possible.This list is not exhaustive:
bending, disassembly, or fracture of implant and components.Loosening of spinal fixation implants may occur due to inadequate initial fixation, latent infection, and/or premature loading, possibly resulting in bone erosion, migration, or pain.
|