Abbott Medical MitraClip G5 Delivery System; mitral valve repair devices Back to Search Results
Model Number CDS0801-XTW
Medical Device Problem Code Incomplete Coaptation (2507)
Health Effect - Clinical Code Mitral Valve Insufficiency/ Regurgitation (4451)
Date of Event 08/14/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2026, a mitraclip procedure was performed to treat mixed mitral regurgitation (mr) with a grade of 4 on a patient with a restricted posterior mitral leaflet (pml) and an anterior leaflet overriding pml.A first clip, a mitrclip xtw (60123a1033) was inserted and deployed on the mitral valve.A second clip, a mitraclip ntw (50915a1041) was inserted and deployed on the mitral valve, reducing mr to trace.On (b)(6) 2026, during pre-procedural imaging for a triclip procedure, it was observed that the previously implanted mitraclip xtw (60123a1033) had detached from the posterior leaflet and remained attached to the anterior leaflet (single leaflet device attachment/slda), causing mr to increase to a grade of 3.On extensive interrogation, it was determined that there were likely some chordae in the clip without full posterior leaflet insertion.The triclip procedure was then completed successfully.No additional information was provided.
 
Additional Manufacturer Narrative
The device remains implanted in patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MitraClip G5 Delivery System
Common Device Name
mitral valve repair devices
Manufacturer (Section D)
Abbott Medical
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
Abbott Medical Northpoint Reg 3020950818
1820 bastian court
westfield IN 46074
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
+1(651)756-5400
MDR Report Key50693697
Report Number2135147-2026-06076
Device Sequence Number9684600
Product Code NKM
UDI-Device Identifier05415067050779
UDI-Public(01)05415067050779(17)270724(10)60123A1033
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P100009
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCDS0801-XTW
Device Catalogue NumberCDS0801-XTW
Device Lot Number60123A1033
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/14/2026
Supplement Date Received by Manufacturer09/24/2026
Initial Report FDA Received Date09/09/2026
Supplement Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/24/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age86 YR
Patient SexMale
-
-