| Model Number |
72400024 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Codes |
Muscle Weakness/Atrophy (1967); Urinary Incontinence (4572)
|
| Date of Event |
08/10/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient with an artificial urinary sphincter (aus) experienced incontinence.Evaluation revealed that the pump had migrated superiorly within the scrotum, the balloon had migrated from the space of retzius into the inguinal canal, and slight urethral atrophy was present.A revision procedure was performed during which the device was explanted and replaced with a new aus.A smaller size cuff was implanted.No further patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Model number/catalog number: 72404127
serial number: n/a
batch/lot number: 1100120228
model/catalog description: control pump fgs iz
unique identifier (udi) #: (b)(4)
model number/catalog number: 72404131
serial number: n/a
batch/lot number: 1100120348
model/catalog description: belt cuff fgs 4.5 cm iz
unique identifier (udi) #: (b)(4).
|
| |
|
Search Alerts/Recalls
|