| Model Number |
1458Q/92 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
08/20/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient presented with a pocket infection.The device pocket was visibly swollen.It was believed that the pocket infection was procedure-related, as it occurred following a device upgrade procedure.To resolve the issue, the left ventricular (lv) lead was explanted.The device and the remaining leads remained implanted.The patient remained in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
The lead was returned and visual examination found no malfunction.A device history record (dhr) review was performed and review of the sterilization records confirmed normal sterilization cycles for the products.The device met specifications prior to leaving abbott manufacturing facilities.The cause of infection could not be traced to the device.
|
| |
|