| Model Number |
1103-161620 |
| Medical Device Problem Code |
Delivered as Unsterile Product (1421)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the proof-of-sterilization stickers on mtp hemiarthroplasty implants were orange in color, rather than red which indicates exposure to sterilant.The implants were not used and the surgeon converted to a different size of the same product line.
|
| |
|
Additional Manufacturer Narrative
|
|
The two devices reported as part of this adverse event are the same size and part of the same production lot.The implants in question were not used and the patient was converted to the next available size.No patient harm has been reported at this time.The medwatch filing will be updated with any new information as appropriate.
|
| |
|